General Description: W&M approved industrial weighing terminal with a 5" (130mm) TFT color display and numeric keypad.
5.1 Calibration Procedure
- Enter the Calibration Menu (passcode protected).
- Zero Calibration: Ensure the scale is empty. Press the ZERO key. The display confirms the zero point.
- Span Calibration: Place a known test weight (e.g., 50% or 100% of capacity) on the scale.
- Enter the exact weight value using the numeric keys.
- Press ENTER to confirm. The unit calculates the slope (mV/V factor).
Present weekly/monthly checklist in a table. systec it3 user manual
- Manual tip: Use distilled water only. Tap water confuses the sensor. To reset the alarm, you usually need to open the front panel (left side) and check the float switch, then hold the "Stop" button for 3 seconds to clear the error.
Practical Use Cases for the SYSTEC IT3 User Manual
Case 1: Setting Up CAN Bus Communication
The manual’s pinout table for the 9-pin D-sub CAN connector is essential. Many users mistakenly swap CAN_H and CAN_L, leading to bus errors. The IT3 uses pin 2 (CAN_L) and pin 7 (CAN_H) – different from older SYSTEC models. The manual also explains the need for a 120 ohm termination resistor on the last node. Enter the Calibration Menu (passcode protected)